A prospective, multicenter, randomized OCT study of early neointimal condition at first and second months after BuMA Supreme stent versus XIENCE stent implantation in high-bleeding-risk coronary artery disease patients: study protocol for a randomized controlled trial

Abstract

Background

Earlier vascular healing after drug-eluting stent (DES) implantation may reduce the incidence of late stent thrombosis and provide theoretical evidence to shorten dual antiplatelet therapy duration in patients with high bleeding risks. The BuMA supreme stent is a newly developed DES-coated with the sirolimus by using the international patent electronic grafted eG™ technology. Previous randomized trials showed that BuMA stents had better stent-strut coverage at 3-month follow-ups, which were evaluated by optical coherence tomography (OCT). However, there have been a limited number of studies that are directly evaluating the extent of neointima formation at the first and second months after stent implantation in high-bleeding-risk patients with coronary artery disease. This clinical trial is designed to demonstrate the non-inferiority of the BuMA supreme stent compared to the XIENCE stent in early neointimal formation.

Methods/design

This is a prospective, multicenter, randomized trial. Forty patients will be assigned into the first-month OCT group, and another 40 patients into the second-month OCT group. The patients in each cohort will be randomized again into two groups in a 1:1 ratio, either being implanted with the BuMA Supreme stent or the Xience V/Prime/Xpedition stent. The primary endpoint is stent-strut neointimal coverage rate (%) at the first and second months, respectively. Secondary endpoints include neointimal hyperplasia area/volume, neointimal hyperplasia thickness, stent-strut malapposition rate, late lumen loss (LLL), restenosis rate, device/lesion/clinical success rate, device-oriented composite endpoints at the first and second months, stent thrombosis and other serious adverse events and bleeding events at follow-up.

Discussion

The results will provide the first accurate imaging evidence on neointimal formation of the BuMA Supreme stent and the Xience stent at 1–2 months post PCI. The result should inspire further exploration and adjustment of DAPT treatments.

Trial registration

ClinicalTrials.gov, ID: NCT02747329. Registered on 21 April 2016. Last updated 17 May 2018.

Peer Review reports

Background

Drug-eluting stents (DES) reduce the rate of restenosis and consequent need for target lesion revascularization (TLR) compared with bare metal stents (BMS). However, impaired endothelialization increases the risk of late and very late stent thrombosis [1]. As a solution, the patient’s postoperative dual anti-platelet therapy (DAPT) treatment increases gradually from 1 month for the BMS to 6–12 months for DES post procedure, or even longer. But extension of DAPT increased the risk of bleeding, which raises another issue of how to shorten the DAPT period after the new DES implantation – which will benefit the patients with relatively high bleeding risk [2].

The BuMA stent (SINOMED, Beijing, China) is a novel, biodegradable, poly-lactic-co-glycolic acid polymer sirolimus-eluting stent (SES), with a thin-strut cobalt-chromium platform (80 μm), thin (200 nm) electro-grafting base between the polymer and stainless-steel stent strut. The BuMA Supreme SES has a pharmacokinetic polymer degradation/drug release profile where the top coat releases the drug relatively rapidly such that sirolimus release from the stents surface is nearly complete by 28 days after implantation [3]. The PIONEER trial showed that the BuMA stent did not meet the criteria for non-inferiority to zotarolimus-eluting stents (ZES) in terms of angiographic in-stent late lumen loss (LLL) at 9-month follow-up [4]. The study also showed that the BuMA SES was superior to the EXCEL SES in endothelialization at 3-month follow-up [5]. However, the earlier (≤ 2-month) endothelialization data are limited. This study can provide accurate imaging evidence on early neointimal formation of the BuMA and Xience stents, and provide evidence for further exploration and adjustment of DAPT treatment.

Methods/design

Study design

This is a prospective, multicenter, randomized clinical trial that should demonstrate the non-inferiority of the BuMA Supreme stent compared with the Xience stent on neointimal healing response evaluated by optical coherence tomography (OCT). In the current trial, the following devices will be used: BuMA Supreme stent – Biodegradable Polymer SES Coated Coronary Stent System manufactured by SINO Medical Sciences Technology Inc. and the Xience V/Prime/Xpedition stent – Durable Polymer everolimus-eluting Stent System manufactured by Abbott Company. A full range of commercially available guiding catheters, balloon catheters, and guidewires will be readily available.

Patient enrollment

Eighty coronary artery disease (CAD) patients with a high bleeding risk will be included from eight centers, the major inclusion and exclusion criteria are summarized in Table 1. Patients will be randomized into two cohorts using a 1:1 ratio, 40 patients will be assigned into a first-month OCT group. The other 40 patients will be assigned into a second-month OCT group. The patients in each cohort will be randomized again into two groups in a 1:1 ratio, either receiving the BuMA Supreme stent or the Xience stent. The process of patient recruitment, enrollment and randomization is summarized in Figs. 1 and 2 shows the Standard Protocol items: Recommendations for Interventional Trials (SPIRIT) Figure for the trial.

Table 1 Major inclusion and exclusion criteria
Fig. 1
figure 1

Design of study. OCT optical coherence tomography, CAD coronary artery disease

Fig. 2
figure 2

Standard Protocol Items: Recommendations for Interventional Trials (SPIRIT) Figure

Randomization

This is a prospective, multicenter, randomized controlled clinical trial; the central randomized method based on the computer system was used in this trial. Patients with provided written, informed consent who meet the “all inclusion and no exclusion” criteria will be enrolled into the study. After the investigator fills in the randomized application form and logs into the randomization website, the computer system will automatically assign random number and corresponding treatment groups to patients according to the patient condition. After obtaining a randomized result, the investigator should record the random number and the corresponding stent information on the randomized application form and the original case history record. Patients will be treated by the system according to the requirement of the randomized groups (the treatment group/the control group). The investigator needs to print a random result list after randomization. The clinical center should archive the randomized list and other related documents (randomized application form) together in the folder. The folder should be archived well in preparation for inspection.

Ethical and informed consent of clinical trials

This trial will be conducted in compliance with the current Helsinki Declaration, Medical Device Clinical Trail Regulations, International Conference on Harmonization (ICH) Good Clinical Practice (GCP) guidelines and relevant national regulations. Prior to initiation of the study, the investigator will submit copies of the protocol, informed consent form and all other relevant documents to the Institutional Review Board (IRB)/Ethics Committees (EC) for its review and approval. A copy of the written IRB/EC approval must be provided to the sponsor (or designee). Any amendments to the protocol, as well as possible associated information and consent form changes, will be submitted to the IRB/EC and written approval obtained prior to implementation. Documentation of serious adverse events in the clinical trials should be submitted to the IRB/EC in time. Eligible patients (and/or the patient’s parents or the patient’s legal guardian, if applicable) should be informed about the trial and have to sign an informed consent prior to the first study procedure. The contents of the trial, purpose of the trial, expected efficacy, possible adverse events and response to adverse events of the procedures and study should be explained to the patient. Informed consents are obtained from the investigators and participants or their legally authorized representatives, and both parties save a copy.

Medication specification

DAPT is recommended for pre-procedure and post-procedure medication in accordance with the “Chinese guideline for percutaneous coronary intervention [7]”; the proposed dose is as follows. In consideration of patients with a high-risk of bleeding, the actual loading dose may vary based on the investigator’s discretion, which will not be deemed as protocol deviation.

For patients who are already on aspirin, a maintenance dose of 100–300 mg/qds will be administered. For patients not previously exposed to aspirin, a loading dose of aspirin 300 mg will be administered as the first dose 2 h before percutaneous coronary intervention (PCI), preferably within 24 h before PCI.

For patients already on clopidogrel, a maintenance dose of 75–150 mg clopidogrel will be administered within 6 h before PCI. For patients who not previously exposed to clopidogrel, a loading dose of 600 mg of clopidogrel will be administered within 6 h before PCI. For patients already on clopidogrel for more than 72 h, no more increased load is needed.

During the procedure, patients will receive appropriate heparin according to actual situation by investigator.

Post the procedure, a subcutaneous low-molecular-weight heparin injection is given if necessary, dual anti-platelet therapy with aspirin and clopidogrel is recommended, 75 mg clopidogrel and 100 mg aspirin once a day. DAPT is recommended for at least 3 months after PCI, a maintenance of DAPT, a reduction in antiplatelet therapy or less antiplatelet treatment time using DAPT is permitted in patients by the investigator according to patient’s specific bleeding and ischemic status.

OCT and QCA evaluation

The OCT and quantitative coronary angiography (QCA) results of this study will be evaluated by an independent core laboratory (PLA general Hospital, Beijing, China) in a blind manner. According to the standard procedure, imaging discs of the baseline procedure, baseline OCT, follow-up angiography and OCT (1 month or 2 months), and any unintended angiography will be sent to the coronary angiography core laboratory to complete the qualitative and quantitative analysis by the experienced image analyst blinded to the treatment protocol. Standard QCA methodology was used including analysis of stent and peri-stent segments defined as spanning 5 mm proximal and distal to the stent edge. Binary restenosis was defined in every segment (proximal, distal and in-stent) as > 50% diameter stenosis at follow-up. LLL was defined as the difference between follow-up and post-procedure minimal lumen diameter.

OCT parameters were calculated and defined as follows [8, 9]: neointimal thickness was measured as the perpendicular distance between the endoluminal surfaces of the neointima to the stent strut. Stent and lumen areas will be measured, and the neointimal hyperplasia area is calculated as the stent area minus the lumen area. The neointimal hyperplasia volume is calculated as the sum of the neointimal hyperplasia areas. The absence of definite neointima over the stent strut was defined as an uncovered stent strut. The distance between the inner surface of the strut reflection and the vessel wall was measured by extending the contours of the walls on the outside of the strut shadow. Stent-strut malapposition was defined as struts with detachment from the vessel wall > 138 μm for the BuMA and > 155 μm for the XIENCE (stent-strut thickness + coating thickness + OCT resolution limit of 20 μm).

Study endpoints and clinical follow-up

The primary endpoint is the stent-strut neointimal coverage rate (%) by OCT assessment at 1 month and 2 months, respectively. Secondary OCT endpoints include neointima hyperplasia area/volume, neointimal hyperplasia thickness and the stent-strut malapposition at 1 or 2 months’ follow-up, respectively. Secondary angiographic endpoints include LLL, restenosis rate of in-stent/in-segment at 1 or 2 months’ follow-up, respectively. Secondary clinical endpoints include device/lesion/clinical success rate, device-oriented composite endpoints and its individual components, other serious adverse (MI, TLR and stent thrombosis) and bleeding events at 1 or 2, 3, 6 months, 1 year and 2 years.

Target lesion failure was defined as the composite of cardiac death, target vessel myocardial infarction (MI), and ischemia-driven TLR. MI was adjudicated according to the third universal definition of MI [10]. Stent thrombosis (definite and probable) was defined according to the Academic Research Consortium (ARC) classification. All clinical events were adjudicated by an independent Clinical Events Committee (CEC). Device success was defined as attainment of < 30% residual stenosis of the target lesion by visual assessment. Lesion success was defined as attainment of < 30% residual stenosis, TIMI 3 flow, and no residual dissection or thrombosis of the target lesion using any percutaneous method. Clinical success was defined as attainment of lesion success of the target lesion and no in-hospital major adverse cardiac event.

Angiography and OCT investigations are conducted at 1 or 2 months post PCI. Clinical follow-up visits or telephone follow-up will be conducted at the first, second, third, sixth month, or 1 to 2 years post PCI.

Sample size of clinical trials

The study’s primary endpoint was stent-strut neointimal coverage rate (%) by OCT assessment at the first or second month. Using a two-sided α of 0.05, 20 subjects were enrolled in each group. It was assumed that 1.5 stents would be implanted in each patient with 200 stent struts (20 mm length) for each stent. Regarding a 20% drop-out rate, 4800 stent struts were estimated to be tested. Assuming a neointimal coverage rate for both the BuMA and Xience stents of 70%, and a non-inferior margin of 3.6% at the first month, stent struts would have at least 97% power to detect that the Supreme group show non-inferiority to that of the Xience group. Assuming a neointimal coverage rate for both BuMA and Xience stents of 85%, and a non-inferior margin of 2.4% at the second month, stent struts would have at least 91% power to detect that the Supreme group shows non-inferiority to that of the Xience group. Since a competition mechanism is introduced into this multicenter study, no single center will enroll more than 40% of the patients.

Statistical analysis

Continuous variables will be described using means and standard deviations or median and range in the case of asymmetric data distribution. Categorical variables will be presented as numbers and frequencies, and comparisons will be made using the chi-square test or Fisher’s exact test as appropriate. All time-to-event outcomes will be summarized using Kaplan-Meier survival estimates and compared among treatment groups using log-rank tests. An array of prespecified and exploratory subgroup analyses of the primary and major secondary endpoints will be performed. Statistical significance will be considered at P < 0.05. A statistical package (SPSS 16.0) will be used for analysis. The individual will be considered as a unit of analysis.

Discussion

PCI is widely used for patients with CAD all over the world. Compared with BMS, DES reduce the rate of restenosis from 20–35% to 10% and reduce the consequent need for TLR. However, DES increases the risk of late stent thrombosis, death and MI [11,12,13]. A previous study [12] reported that the incidence of definite stent thrombosis was increased by 0.53% (CI 0.44–0.64) per year, and the cumulative rate of definite stent thrombosis was 3.3%, combined incidence of definite stent thrombosis and the probable stent thrombosis was 5.7% in 4 years. At present, some studies have regarded endothelialization as one of the important factors that may reduce late thrombosis and then reduce cardiovascular events; therefore, research on DES has focused on how to promote the endothelialization. The ideal DES is to maximize the inhibition of smooth-muscle-cell proliferation and migration, minimize changes of structure and functional recovery of endothelial cell, and promote faster healing of endothelium [14, 15]. The second generation of drug-eluting stents (DES) has emerged with more biocompatible durable polymer coatings on thin-strut stents that have shown better safety than first-generation DES [16, 17]. The Biodegradable Polymer (BP)-DES, BuMA Supreme stent, was developed to prevent polymer-related adverse effects [18].

DAPT treatment increases from 1 month of the BMS to 6–12 months of DES post procedure, or even longer, which increases the risk of bleeding. Research on how to shorten the DAPT period after the new DES implantation will benefit relatively high-bleeding-risk patients. Studies showed that one out of seven to eight CAD PCI patients has a relatively high risk of bleeding, so clinicians need to evaluate bleeding and ischemic risk after DES, and adjust their DAPT treatment to obtain better clinical results [2].

In the recent 10 years, OCT has been widely used as an intravascular imaging technology to determine the morphology of the implanted stents, whereby high-resolution intravascular images can be obtained quickly – especially accurate results of stent-strut coverage in the vessel [19, 20]. For example, Ishigami K et al. observed stent-strut coverage of the first-generation sirolimus eluting stent (SES) at 9 months, 9–24 months and 25 months, respectively, and they found that only 17.6% of the SES was completely covered at 25 months. Delayed neointimal coverage was observed in patients with small blood vessels, complex lesions, high-lipid and high-calcium lesion or diabetics [21]. OCT was a reliable method for assessing the neointimal formation of stent strut post PCI, and it provided a good match to the resemblance of “optical biopsy.”

The research and development of stents is focused on the materials and coating composition of stents in China [22]. OCT is rarely used to evaluate vessel healing after new stent implantation. We have evaluated new BuMA DES strut coverage at 9 months by OCT in 2010, which showed that stent-strut coverage rate in BuMA stent was 94.3%, which was superior to that of the Endeavor stent [23]. The PANDA-III study with 2348 samples showed a statistically significant difference in the incidence of definite and probable stent thrombosis (0.7% vs. 1.4%) at 1 year, that is less stent thrombosis than occurred in the BuMA stent [24]. Moreover, in OCT follow-up at 3 months, the percentage of stent-strut coverage was significantly higher in the BuMA vs. the EXCEL group (strut level: 94.2% vs. 90.0%, P < 0.01) [5]. It showed that the design concept and technology of BuMA stent is good to promote early and complete healing of the neointima at a corresponding time after PCI at 3 months. The LEADERS FREE study also used large samples to compare the efficacy and safety of implantation of BMS and new DES in patients with CAD; 1-year follow-up showed that the efficacy and safety of the new DES group was superior to that of the BMS group [6, 25]. The OCT study of neointimal coverage has been reduced from 9 months to 3 months at present [5]; one purpose of this study is whether we could learn more about stent-strut coverage rate at 2 months and 1 month, which will help to conduct further study of an adjusted postoperative DAPT regimen. The evidence may provide clinicians and patients with more precise imaging evidence in DAPT regimens, especially more benefit for patients with a relatively high bleeding risk.

For BuMA stents that used international-patent, electronic-grafted (eG) technology, common problems, such as stent surface cracking and residue after drug release, are solved successfully, leading to rapid and complete endometrial healing and inhibition of late stent thrombosis. This feature is more conducive to helping patients with a relatively high risk of bleeding to adjust their DAPT treatment regimen. Based on the design of the previous generation of BuMA stents, the stent metal platform of the BuMA Supreme changed from 316 L stainless steel to L605 cobalt chromium – with thinner stent strut and better stent effects in contrast imaging and flexibility. Meanwhile, the thinner stent strut may further benefit the endothelium healing, which will bring the patients more safety benefits.

In conclusion, this is the first study to compare the strut coverage of the XIENCE stent with that of the BuMA Supreme sirolimus-eluting cobalt-chromium stent, which has a shorter drug elution on optical coherence tomography (OCT) 1 or 2 months after implantation; and it will give us accurate imaging evidence on neointimal formation of the BuMA Supreme stent and the Xience stent at 1–2 months post PCI. The result should inspire further exploration and adjustment of DAPT treatments Additional file 1.

Abbreviations

ARC:

Academic Research Consortium

BMS:

Bare metal stent

CAD:

Coronary artery disease

DAPT:

Dual anti-platelet therapy

DES:

Drug-eluting stents

LLL:

Late lumen loss

LVEF:

Left ventricular ejection fraction

MLD:

Minimal lumen diameter

OCT:

Optical coherence tomography

PCI:

Percutaneous coronary intervention

PLGA:

Poly-lactic-co-glycolic acid

PTCA:

Percutaneous transluminal coronary angioplasty

QCA:

Quantitative coronary angiography

RVD:

Reference vessel diameter

SES:

Sirolimus-eluting stent

TLF:

Target-lesion failure

TLR:

Target-lesion revascularization

ZES:

Zotarolimus-eluting stents

References

  1. Joner M, Finn AV, Farb A, Mont EK, Kolodgie FD, Ladich E, Kutys R, Skorija K, Gold HK, Virmani R. Pathology of drug-eluting stents in humans: delayed healing and late thrombotic risk. J Am Coll Cardiol. 2006;48(1):193–202.

    Article  Google Scholar 

  2. Urban P, Abizaid A, Chevalier B, Greene S, Meredith I, Morice MC, Pocock S. Rationale and design of the LEADERS FREE trial: a randomized double-blind comparison of the BioFreedom drug-coated stent vs the Gazelle bare metal stent in patients at high bleeding risk using a short (1 month) course of dual antiplatelet therapy. Am Heart J. 2013;165(5):704–9.

    Article  CAS  Google Scholar 

  3. Chen M, Wang XG, Zheng B, Peng HY, Zhang XY, Zhang B, Huo Y. Investigation of long-term implantation of BuMA stent in a porcine coronary model. Chin Med J. 2012;125(22):4083–7.

    PubMed  Google Scholar 

  4. von Birgelen C, Asano T, Amoroso G, Aminian A, Brugaletta S, Vrolix M, Hernandez-Antolín R, van de Harst P, Iñiguez A, Janssens L, et al. First-in-man randomised comparison of the BuMA Supreme biodegradable polymer sirolimus-eluting stent versus a durable polymer zotarolimus-eluting coronary stent: the PIONEER trial. EuroIntervention. 2018;13(17):2026–35.

    Article  Google Scholar 

  5. Qian J, Zhang YJ, Xu B, Yang YJ, Yan HB, Sun ZW, Zhao YL, Tang YD, Gao Z, Chen J, et al. Optical coherence tomography assessment of a PLGA-polymer with electro-grafting base layer versus a PLA-polymer sirolimus-eluting stent at three-month follow-up: the BuMA-OCT randomised trial. EuroIntervention. 2014;10(7):806–14.

    PubMed  Google Scholar 

  6. Urban P1, Meredith IT, Abizaid A, Pocock SJ, Carrié D, Naber C, Lipiecki J, Richardt G, Iñiguez A, Brunel P, et al. Polymer-free drug-coated coronary stents in patients at high bleeding risk. N Engl J Med. 2015;373(21):2038–47.

    Article  CAS  Google Scholar 

  7. Section of Interventional Cardiology of Chinese Society of Cardiology of Chinese Medical Association; Specialty Committee on Prevention and Treatment of Thrombosis of Chinese College of Cardiovascular Physicians; Editorial Board of Chinese Journal of Cardiology. Chinese guideline for percutaneous coronary intervention. Zhonghua xin xue guan bing za zhi. 2016;44(5):382–400.

    Google Scholar 

  8. Tearney GJ, Regar E, Akasaka T, Adriaenssens T, Barlis P, Bezerra HG, Bouma B, Bruining N, Cho JM, Chowdhary S, et al. Consensus standards for acquisition, measurement, and reporting of intravascular optical coherence tomography studies: a report from the International Working Group for Intravascular Optical Coherence Tomography Standardization and Validation. J Am Coll Cardiol. 2012;59(12):1058–72.

    Article  Google Scholar 

  9. F P, Guagliumi G, Mintz GS, Costa M, Regar E, Akasaka T, Barlis P, Tearney GJ, Jang IK, Arbustini E, et al. Expert review document part 2: methodology, terminology and clinical applications of optical coherence tomography for the assessment of interventional procedures. Eur Heart J. 2012;33(20):2513–20.

    Article  Google Scholar 

  10. Thygesen K, Alpert JS, Jaffe AS, Simoons ML, Chaitman BR, White HD, Joint ESC/ACCF/AHA/WHF Task Force for Universal Definition of Myocardial Infarction; Authors/Task Force Members Chairpersons, Thygesen K, Alpert JS, White HD, et al. Third universal definition of myocardial infarction. J Am Coll Cardiol. 2012;60(16):1581–98.

    Article  Google Scholar 

  11. Daemen J, Wenaweser P, Tsuchida K, Abrecht L, Vaina S, Morger C, Kukreja N, Jüni P, Sianos G, Hellige G, et al. Early and late coronary stent thrombosis of sirolimus-eluting and paclitaxel-eluting stents in routine clinical practice: data from a large two-institutional cohort study. Lancet. 2007;369(9562):667–8.

    Article  CAS  Google Scholar 

  12. Wenaweser P, Daemen J, Zwahlen M, van Domburg R, Jüni P, Vaina S, Hellige G, Tsuchida K, Morger C, Boersma E, et al. Incidence and correlates of drug-eluting stent thrombosis in routine clinical practice. 4-year results from a large 2-institutional cohort study. J Am Coll Cardiol. 2008;52(14):1134–40.

    Article  CAS  Google Scholar 

  13. Windecker S, Meier B. Late coronary stent thrombosis. Circulation. 2007;116(17):1952–65.

    Article  Google Scholar 

  14. Pfisterer ME. Late stent thrombosis after drug-eluting stent implantation for acute myocardial infarction: a new red flag is raised. Circulation. 2008;118(11):1117–9.

    Article  Google Scholar 

  15. Benezet-Mazuecos J, Ibanez B, Badimon JJ. Dual antiplatelet therapy and drug eluting stents: a marriage of convenience. Thromb J. 2007;5:15.

    Article  Google Scholar 

  16. Stone GW, Rizvi A, Sudhir K, Newman W, Applegate RJ, Cannon LA, Maddux JT, Cutlip DE, Simonton CA, Sood P, Kereiakes DJ, SPIRIT IV Investigators. Randomized comparison of everolimus- and paclitaxel-eluting stents. 2-year follow-up from the SPIRIT (Clinical Evaluation of the XIENCE V Everolimus Eluting Coronary Stent System) IV trial. J Am Coll Cardiol. 2011;58(1):19–25.

    Article  Google Scholar 

  17. Räber L, Magro M, Stefanini GG, Kalesan B, vanDomburg RT, Onuma Y, Wenaweser P, Daemen J, Meier B, Jüni P, et al. Very late coronary stent thrombosis of a newer-generation everolimus-eluting stent compared with earlygeneration drug-eluting stents: a prospective cohort study. Circulation. 2012;125(9):1110–21.

    Article  Google Scholar 

  18. Stefanini GG, Byrne RA, Serruys PW, de Waha A, Meier B, Massberg S, Jüni P, Schömig A, Windecker S, Kastrati A. Biodegradable polymer drug-eluting stents reduce the risk of stent thrombosis at 4 years in patients undergoing percutaneous coronary intervention: a pooled analysis of individual patient data from the ISAR-TEST 3, ISAR-TEST 4, and LEADERS randomized trials. Eur Heart J. 2012;33(10):1214–22.

    Article  CAS  Google Scholar 

  19. Gutiérrez-Chico JL, Regar E, Nüesch E, Okamura T, Wykrzykowska J, di Mario C, Windecker S, van Es GA, Gobbens P, Jüni P, et al. Delayed coverage in malapposed and side-branch struts with respect to well-apposed struts in drug-eluting stents: in vivo assessment with optical coherence tomography. Circulation. 2011;124(5):612–23.

    Article  Google Scholar 

  20. Gomez-Lara J, Radu M, Brugaletta S, Farooq V, Diletti R, Onuma Y, Windecker S, Thuesen L, McClean D, Koolen J, et al. Serial analysis of the malapposed and uncovered struts of the new generation of everolimus-eluting bioresorbable scaffold with optical coherence tomography. JACC Cardiovasc Interv. 2011;4(9):992–1001.

    Article  Google Scholar 

  21. Ishigami K, Uemura S, Morikawa Y, Soeda T, Okayama S, Nishida T, Takemoto Y, Onoue K, Somekawa S, Takeda Y, et al. Long-term follow-up of neointimal coverage of sirolimus-eluting stents—evaluation with optical coherence tomography. Circ J. 2009;73(12):2300–7.

    Article  Google Scholar 

  22. Qian J, Xu B, Lansky AJ, Yang YJ, Qiao SB, Wu YJ, Chen J, Hu FH, Yang WX, Mintz GS, et al. First report of a novel abluminal groove filled biodegradable polymer rapamycin-eluting stent in de novo coronary artery disease: results of the first in man FIREHAWK trial. Chin Med J. 2012;125(6):970–6.

    CAS  PubMed  Google Scholar 

  23. Liu CF, Chen YD, Chen L, Sun ZJ, Gai LY, Liu HB, Tian F, Bai QC, Guo K. Optical coherence tomography for evaluation of neointimal proliferation after placement of a new drug eluting stent. Nan Fang Yi Ke Da Xue Xue Bao. 2010;30(5):1063–5.

    PubMed  Google Scholar 

  24. Xu B, Gao R, Yang Y, Cao X, Qin L, Li Y, Li Z, Li X, Lin H, Guo Y, et al. Biodegradable polymer-based sirolimus-eluting stents with differing elution and absorption kinetics: The PANDA III Trial. J Am Coll Cardiol. 2016;67(19):2249–58.

    Article  CAS  Google Scholar 

  25. Serruys PW, Onuma Y, Garg S, Vranckx P, De Bruyne B, Morice MC, Colombo A, Macaya C, Richardt G, Fajadet J, et al. 5-year clinical outcomes of the ARTS II (Arterial Revascularization Therapies Study II) of the sirolimus-eluting stent in the treatment of patients with multivessel de novo coronary artery lesions. J Am Coll Cardiol. 2010;55(11):1093–101.

    Article  Google Scholar 

Download references

Acknowledgements

We are grateful to the numerous colleagues who are providing clinical and research support in treating and following the patients included in this trial. This work was confirmed as Scientific Innovation Research supported by the Chinese PLA General Hospital. The funding body had no role in designing the study, nor will it be involved in the collection, analysis, or interpretation of data.

Funding

This study was funded by SINOMED, Tianjin, China.

Availability of data and materials

All data will be archived in individual patient medical records under the responsibility of the Chinese PLA General Hospital in Beijing China.

The data set generated and analyzed during the current study will be available from the corresponding author on reasonable request.

Trial status

The trial is currently in the recruitment phase.

Author information

Authors and Affiliations

Authors

Contributions

YDC, the principal investigator (PI) of this study, was involved in the conception and design of the study. BL and QHJ contributed to the drafting and writing of the manuscript. CQW, BS, XS, GSF, YQW, XCZ and ZYY all managed the project at each hospital or medical center. All authors have read and approved the submission of this manuscript.

Corresponding author

Correspondence to Yun Dai Chen.

Ethics declarations

Author’s information

Not applicable.

Ethics approval and consent to participate

This study has been approved by the Ethics Committee Board of Chinese PLA General Hospital, Beijing, China, which was accredited by the SIDCER/FERCAP Survey in 2012. Written informed consent was obtained from each patient/participant for publication of their individual details and accompanying images in this manuscript. The consent form is held by the authors/by the authors’ institution/in the patients’ clinical notes.

Consent for publication

Not applicable.

Competing interests

The authors declare that they have no competing interests.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Additional file

Additional file 1:

SPIRIT 2013 Checklist: recommended items to address in a clinical trial protocol and related documents*. (DOC 130 kb)

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Li, B., Jin, Q.H., Chen, Y.D. et al. A prospective, multicenter, randomized OCT study of early neointimal condition at first and second months after BuMA Supreme stent versus XIENCE stent implantation in high-bleeding-risk coronary artery disease patients: study protocol for a randomized controlled trial. Trials 20, 335 (2019). https://doi.org/10.1186/s13063-019-3361-0

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s13063-019-3361-0

Keywords